  • Dust particle counter INS-DPC01

  • Online particle counter can guarantee 24 *7 Uninterrupted work with a long life span of ten years (Note: Calibration and calibration must be carried out every year ) 。The online microparticle counter includes a laser sensor and isopower sampling probe.Online dust particle counter is used for detection of clean rooms and other clean environments. The counter is based on the principle of light scattering as the theoretical basis and uses photoelectric detection technology to measure and count the particle size and number of dust particles.The system has the advantages of good stability and accurate testing.


Product Manual

Product Specification Book

  • Product Specification Book

    Product Name:

    Online dust particle counter

    Product model:

    INS-DPC01

    Version:

    V1.0

    Formulator:

    He Genwen

    Reviewer:

    Yuan Chao

     

    1. Product Overview

    Online particle counter can guarantee 24 *7  Uninterrupted work with a long life span of ten years  (Note: The calibration must be carried out every year under the premise of ) 。The online microparticle counter includes a laser sensor and isopower sampling probe.

    Online dust particle counter is used for detection of clean rooms and other clean environments. The counter is based on the principle of light scattering as the theoretical basis and uses photoelectric detection technology to measure and count the particle size and number of dust particles.The system has the advantages of good stability and accurate testing.

     

    2. Product Features

      RS485  Serial data transmission;

      LCD touch screen display, high-end and atmospheric;

      Small sensors monitor space microparticles;

      Good system stability and accurate test;

      Has ten years of long-lasting service life  (Note: Calibration and calibration must be carried out every year ) ;

      Provides 6 particle size channels: 0.3  μm、0.5 μm、1.0 μm、3.0 μm、5.0μm  and 10.0 μm。

     

    3. Interface definition

    Pin

    name

    describe

    Remark

    1

    VCC

    The power supply of the device is

    DC24V

    2

    GND

    Equipment grounding

    ——

    3

    NC

    ——

    ——

    4

    485B

    Communication pins

    (RS485+)

    5

    485A

    Communication pins

    (RS485- )

     

    4. Performance parameter table

    1 、Parameter table

    Particle size range:

    0.3– 10 μm

    Channel size:

    Standard: 0.5, 5.0,  μm

    Counting performance:

    50% @ 0.5 μm; 100% for particles > 0.75 μm (per JIS)

    Flow rate:

    0.1 CFM (2.83 LPM)

     

    Laser source:

    Laser Diode , Laser diode

    Zero count:

    <1 count / 5 minutes (per JIS) ,<1  count /5  minute  (Every  JIS)

    Concentration control:

    5%At 2,000,000 pills /Cubic ft

    Vacuum requirements:

    Built-in vacuum pump

    Communication mode:

    RS-485 MODBUS   protocol

    LED   Indicator light:

    Power, operation, flow, alarm

    Connector:

    MG 15EDGWB-3.81-5P

    software:

    Special version for the pharmaceutical industry NETSCADA

    shell:

    Stainless Steel , Stainless Steel 304

    power supply:

    DC 24 V

    size:

    13 cm(long ) x 12 cm(Width ) x 8cm(high )

    weight:

    0.5  Kilogram

    power

    5W

     

    2 、Environmental requirements:

    Operation: Temperature 50 °F   To 104 °F (10°C   To 40 °C) /  Humidity 20%  to 95%   No condensation.Storage: Temperature 14 °F   By 122 °F (- 10°C   To 50 °C) /  Humidity 98% No condensation.

    3 、Accessories: isopower sampling probe, probe hat, pipeline, etc.

    4 、Calibration specifications

    1 、conform to  JJF1190-2008   Dust particle counter calibration specification requirements:

    2 、 Self-purification time ≤10min

    3 、Repeatability ≤10%FS

    4 、Particle size distribution error ≤±30%

    5 、Particle concentration display error ≤±30%FS

    6 、Sampling flow error: 2.83 L/min  ≤±5%

    7 、HLD780   Long-life laser diode working time ≥30000  Hour.

     

    V. Communication Protocol

    1、Register description  (The base address of this address is 1 )

    Modbus  address

    read /Write

    Data type

    describe

    1

    Read, write

    Integer

    The current address of the instrument  (1-250,Factory default 1 ),Broadcast address 252

    2

    Read, write

    Integer

    Communication baud rate: 0-9600 (Factory default ) ,1-19200, 2-38400

    3

    Read, write

    Integer

    Communication check digit: 0-No check  (Factory default ) ,1-Even check, 2-odd check

    4

    Read, write

    Integer

    Start and stop instrument  (Set 1, start the instrument;0: Stop the instrument ) Or read the operating status of the instrument (1:Start, 0: Stop )

    5

    Read, write

    Integer

    Counting delay (Second ) 0 100 Second

    6

    Read, write

    Integer

    Counting cycle (Second ) 2


    7

    read

    Integer

    spare

    8

    Read, write

    Integer

    Run the buzzer: Count the full cycle whether the buzzer sounds:

     




    0-No sound  (Factory default ) ,1-ring

    9

    Read, write

    Integer

    Alarm buzzer: When the sampling period data alarm is alarmed, whether the buzzer sounds: 0-Not  (Factory default ) ,1-ring

    10

    read

    Integer

    When the counter is currently counted (Second )

    11-12

    read

    Integer

    spare

    13

    read

    Integer

    0.3um   The cumulative data is 16 bits lower

    14

    read

    Integer

    0.3um   The cumulative data is 16 digits higher

    15

    read

    Integer

    0.5um   The cumulative data is 16 bits lower

    16

    read

    Integer

    0.5um   The cumulative data is 16 digits higher

    17

    read

    Integer

    1.0um   The cumulative data is 16 bits lower

    18

    read

    Integer

    1.0um   The cumulative data is 16 digits higher

    19

    read

    Integer

    3.0um   The cumulative data is 16 bits lower

    20

    read

    Integer

    3.0um   The cumulative data is 16 digits higher

    21

    read

    Integer

    5.0um   The cumulative data is 16 bits lower

    22

    read

    Integer

    5.0um   The cumulative data is 16 digits higher

    23

    read

    Integer

    10.0um  The cumulative data is 16 bits lower

    24

    read

    Integer

    10.0um  The cumulative data is 16 digits higher

    25

    read

    Integer

    0.3um   Periodic data is 16 bits lower

    26

    read

    Integer

    0.3um   Cycle data is 16 bits higher

    27

    read

    Integer

    0.5um   Periodic data is 16 bits lower

    28

    read

    Integer

    0.5um   Cycle data is 16 bits higher

    29

    read

    Integer

    1.0um   Periodic data is 16 bits lower

    30

    read

    Integer

    1.0um   Cycle data is 16 bits higher

    31

    read

    Integer

    3.0um   Periodic data is 16 bits lower

    32

    read

    Integer

    3.0um   Cycle data is 16 bits higher

    33

    read

    Integer

    5.0um   Periodic data is 16 bits lower

    34

    read

    Integer

    5.0um   Cycle data is 16 bits higher

    35

    read

    Integer

    10.0um  Periodic data is 16 bits lower

    36

    read

    Integer

    10.0um  Cycle data is 16 bits higher

    37

    read

    Integer

    Current traffic data /100=Current traffic

    38

    read

    Integer

    temperature (With 1 decimal place, the value 100 means 10.0 ℃)

    39

    read

    Integer

    humidity (With 1 decimal place, the value 100 means 10.0% )

    40

    read

    Integer

    Atmospheric pressure (With 2 decimal places, the value 10000 means 100.00 Kpa)

    41

    Read, write

    Integer

    0.3um  Correction coefficient (%):0~10000%,Default 100%

    42

    Read, write

    Integer

    0.5um  Correction coefficient (%):0~10000%,Default 100%

    43

    Read, write

    Integer

    1.0um  Correction coefficient (%):0~10000%,Default 100%

    44

    Read, write

    Integer

    3.0um  Correction coefficient (%):0~10000%,Default 100%

     

    45

    Read, write

    Integer

    5.0um  Correction coefficient (%):0~10000%,Default 100%

    46

    Read, write

    Integer

    10.0um  Correction coefficient (%):0~10000%,Default 100%

    47

    Read, write

    Integer

    Temperature correction value (℃):-100~100(-10.0~ 10.0℃)

    48

    Read, write

    Integer

    Humidity correction value (%):-100~100(-10.0~ 10.0%)

    49

    Read, write

    Integer

    Atmospheric pressure correction value

    (kPa):-1000~1000(-10.00~ 10.00kPa)

    50

    read

    Integer

    spare

    51

    Read, write

    Integer

    0.3um  Alarm enabled

    0:Don't call the police  (Factory default ) ,1:Call the police

    52

    read

    Integer

    0.3um  Alarm status

    0:No alarm, 1: Call the alarm

    53

    Read, write

    Integer

    0.3um   Alarm value is 16 digits lower

    54

    Read, write

    Integer

    0.3um   Alarm value is 16 digits higher

    55

    Read, write

    Integer

    0.5um  Alarm enabled

    0:Don't call the police  (Factory default ) ,1:Call the police

    56

    read

    Integer

    0.5um  Alarm status

    0:No alarm, 1: Call the alarm

    57

    Read, write

    Integer

    0.5um   Alarm value is 16 digits lower

    58

    Read, write

    Integer

    0.5um   Alarm value is 16 digits higher

    59

    Read, write

    Integer

    1.0um  Alarm enabled

    0:Don't call the police  (Factory default ) ,1:Call the police

    60

    read

    Integer

    1.0um  Alarm status

    0:No alarm, 1: Call the alarm

    61

    Read, write

    Integer

    1.0um   Alarm value is 16 digits lower

    62

    Read, write

    Integer

    1.0um   Alarm value is 16 digits higher

    63

    Read, write

    Integer

    3.0um  Alarm enabled

    0:Don't call the police  (Factory default ) ,1:Call the police

    64

    read

    Integer

    3.0um  Alarm status

    0:No alarm, 1: Call the alarm

    65

    Read, write

    Integer

    3.0um   Alarm value: 16 digits lower

    66

    Read, write

    Integer

    3.0um   Alarm value: 16 digits higher

    67

    Read, write

    Integer

    5.0um  Alarm enabled

    0:Don't call the police  (Factory default ) ,1:Call the police

    68

    read

    Integer

    5.0um  Alarm status

    0:No alarm, 1: Call the alarm

    69

    Read, write

    Integer

    5.0um   Alarm value: 16 digits lower

    70

    Read, write

    Integer

    5.0um   Alarm value: 16 digits higher

    71

    Read, write

    Integer

    10.0um  Alarm enabled




    0:Don't call the police  (Factory default ) ,1:Call the police

    72

    read

    Integer

    10.0um  Alarm status

    0:No alarm, 1: Call the alarm

    73

    Read, write

    Integer

    10.0um   Alarm value is 16 digits lower

    74

    Read, write

    Integer

    10.0um   Alarm value is 16 digits higher

    75

    Read, write

    Integer

    Temperature alarm enable

    0:Don't call the police  (Factory default ) ,1:Call the police

    76

    read

    Integer

    Temperature alarm status

    0:No alarm, 1: Call the alarm

    77

    Read, write

    Integer

    Temperature Alarm value is 16 bits lower

    78

    Read, write

    Integer

    Temperature Alarm value is 16 digits higher

    79

    Read, write

    Integer

    Humidity alarm enable

    0:Don't call the police  (Factory default ) ,1:Call the police

    80

    read

    Integer

    Temperature alarm status

    0:No alarm, 1: Call the alarm

    81

    Read, write

    Integer

    Humidity Alarm value is 16 digits lower

    82

    Read, write

    Integer

    Humidity Alarm value is 16 digits higher

    also: Modbus  Data bit 8 bits, stop bit 1 bit

    Special notation: Address 252 is used for broadcasting addresses, as long as it is used in several subsequent situations

    (1)   There is only one instrument on the entire bus

    At this time, you can set the communication address of the current instrument

    (2)   Start or stop all instruments on the entire bus

    If there are multiple instruments on the bus, set the address with 252 address on the bus  (0 Number register ) ,This will cause the addresses of all instruments on the bus to be the same, and the communication will not correctly identify a instrument in the future, causing the entire system to fail.

    2、Meter supported  MODBUS  instruction

     

     

    Function code

    effect

    illustrate

    03

    Read hold register  (16 Bit Register )

    Read one or more hold registers

    06

    Write a single 16-bit register

    Set the current value of a hold register

    16

    Write multiple 16-bit registers

    Set one or more hold registers

     

    6. Things to note

    1、The product is equipped with a laser emitter. Private repairs may cause the operator to be accidentally exposed to laser radiation, and should be repaired by specialized personnel from the manufacturer;

    2、It is prohibited to open the product shell;

    3、Products are not directly in contact with the monitoring environment  (For example, installed in the equipment chassis ) ,In order to accurately sample and measure the results, please connect the air inlet to the external probe of the chassis, and the length of the connecting hose between the external probe should be controlled at 30 cm  Within;

    4、The external probe of the chassis needs to be windproof, rough filtration and waterproof.


  • Home
  • Product Display
  • Qualification certificate
  • Application News